IL1400 XRL140T254
Hand-held, low profile, narrowband germicidal system.
|
ILT1400A or ILT1400BL PHOTOMETER
|
|||
|
![]() |
||
|
DETECTOR: XRL140T254
|
|||
![]() |
|||
|
LOW-PROFILE, HIGH-INTENSITY GERMICIDAL DETECTOR Configured specifically to measure germicidal effectiveness at 254 nm from a wide array of high intensity, germicidal sources. Measurement Range: 3.33e-7 to 1.17e+0 Watts/cm2 Spectral Range: 249 - 259 nm, Center wavelength at 254 nm. Order part numbers:ILT1400A (meter) or ILT1400BL (backlit meter), XRL140T254 (detector) Applications:This detector has a narrow bandwidth with a center wavelength at 254 nm facilitating the measurement of germicidal effectiveness at 254 nm from high intensity germicidal UV sources such as high-pressure or medium-pressure mercury lamps, xenon arc lamps, mercury arc, or mercury-xenon arc lamps. These sources produce a broad spectral UV output with more intense mercury lines than the output at 254 nm which will cause artificially high readings from detectors with a broader response. This detector is built into our low-profile XR housing for use in low clearance applications. |
![]() |
||
|
FILTERS AND INPUT OPTICS ARE NOT REMOVABLE IN THE XRL DETECTORS. |
![]() |
||