Germicidal Light Measurement System - Model ILT1700 SED240/W
Bench-top research germicidal light measurement system
|
ILT1700 RESEARCH RADIOMETER / PHOTOMETER
|
|||
|
![]() |
||
|
DETECTOR: SED240/W
|
|||
![]() |
|||
|
LOW INTENSITY, LOW PRESSURE MERCURY DETECTOR Configured to measure low intensity, low-pressure mercury lamps only. Measurement Range: 1.00e-9 to 5.00e-4 Watts/cm2 UV Actinic Hazard Spectral Range: 185 – 320 nm, calibration wavelength at 254 nm Dimensions: 60 mm x 42 mm diameter Order part numbers: ILT1700 (meter), SED240/W (detector) Applications:This detector has a peak spectral response at 254 nm and is designed to measure low-intensity, low-pressure mercury sources only. Since low-pressure mercury lamps produce better than 95% of their energy at 254 nm, they are easily measured by this detector without additional filtration. |
![]() |
||
|
FILTERS AND INPUT OPTICS ARE REMOVABLE AND MAY BE ORDERED SEPARATELY FROM DETECTOR ASSEMBLY TO ALLOW FILTER AND INPUT OPTIC SWAPPING. |
![]() |
||