ILT5000 Research Radiance Measurement System

ILT Research Radiance Measurement System

ILT5000, SED033/F/R

Part Number: ILT5000, SED033/F/R US
$3,603.00

Research Radiance Measurement System

  • 10+ Decade Dynamic Range of Optical Analysis
  • Pico-ammeter: 10fA to 1mA (calibrated from 100fA - 1mA)
  • Wireless Communication, Built-in
  • User controlled or Auto: Range, Zero/Dark, Sample Rate
  • 4-20 mA Output
  • 15 pin D Sub and SMA sensor/input ports
  • “Set it & Forget It” Remote Data Logging
  • Power: USB or Built-in Re-Chargeable Battery Pack
  • Multi-System Simultaneous Continuous Monitoring
  • Includes DataLight II Software and Labview Sample Code
  • NIST-Traceable/ISO/IEC 17025:2017 Accredited Calibration
  • Sample Rate Up to 100 Hz, Programmable
  • Custom & OEM Inquiries Welcome


< Back to Audience Scanning and Laser Safety Product Table

< Back to LED Testing & Measurement Product Table

033fr

SED033/F/R Narrow field of view, radiance probe. Field of view: 1.5 deg. Designed for average measurement of radiance over a target area greater than 30 mm diameter plus 0.75 degree half angle.

Dynamic Range: 2e-9 to 1e+0 W/cm²/sr
Measurement Range: 400-1064 nm
Dimensions: 87 x 42 mm dia.
Ordering part#(s): ILT5000*, SED033/F/R
* Software Included / Tablet PC Sold Separately
 

SED-SEL detector

 

 

 

 

 

 

 

 

 

Filters and input optics are removable and may be ordered separately from the detector assembly to allow filter and input optic swapping.

 


Note: For simultaneous multi-sensor data logging, ILT also offers the ILT6000 Rack-Mount Radiometer System with the same specifications as the ILT5000 in a 2-10 unit, rack mountable housing.

 

 

 

 

 

 

 

 

 

 

 

 

For the latest meter specifications, data sheet, installation guide, tech notes and software, visit the ILT5000 page here.