ILT1700 Photometer, SED033/Y/H Detector (< 25') Flash/Steady

ILT1700 illuminance measurement system

View ILT5000 Replacement


*** DISCONTINUED ***

The ILT1700 has been replaced the ILT5000 Radiometer which includes complimentary Datalight III software, faster measurement speeds, wireless data transmission, iPhone/iPad application, internal data storage, a broader measurement range and more…

 

ILT5000 Feastures

Bench-Top Research Beam Candela/lluminance Measurement System

  • 10+ Decade Dynamic Range of Optical Analysis
  • Pico-ammeter: 10fA to 1mA (calibrated from 100fA - 1mA)
  • Wireless Communication, Built-in
  • User controlled or Auto: Range, Zero/Dark, Sample Rate
  • 4-20 mA Output
  • 15 pin D Sub and SMA sensor/input ports
  • “Set it & Forget It” Remote Data Logging
  • Power: USB or Built-in Re-Chargeable Battery Pack
  • Multi-System Simultaneous Continuous Monitoring
  • Includes DataLight II Software and Labview Sample Code
  • NIST-Traceable/ISO17025 Accredited Calibration
  • Sample Rate Up to 100 Hz, Programmable
  • Made in the USA
  • Custom & OEM Inquiries Welcome

ILT1700 Specifications

Bench-Top Research Beam Candela/Illuminance Measurement System

  • Industry Standard Research Radiometer
  • Direct Reading in Scientific Notation
  • Best Price Performance
  • Unsurpassed Reliability for 15 years
  • Dynamic Auto Ranging – no lost samples in pulsed applications
  • 10 Decade Dynamic Range
  • 0.1 Percent Linearity
  • 40 Decade Integration Range
  • Light Integration up to 18 Years
  • 10 Calibration Factors in Memory
  • Made in the USA

 

SED033/Y/H curve

SED033/Y/H Illuminance sensor designed for beam candela measurements (steady source only) at a distance of less than 25 feet. Includes the Y filter which was redesigned using proprietary composite filter design software to match the CIE V(?) Photopic curve within 1% total area error.

Dynamic Range: 2.02e-4 to 2.02e+5 lux plus the factor for cd of a continuous source at any distance, at any distance LESS than 7.62 meters (25 feet). Use this ILT1700 system for distances GREATER than 7.62 meters (25 feet).
 

Measurement Range: 400-700 nm
Dimensions: 90 x 42 mm dia.
Ordering part#(s): ILT1700, SED033/Y/H

 

SED033 filter

 

beam intensity

 

 

 

 

 

 

 

FILTERS AND INPUT OPTICS ARE REMOVABLE AND MAY BE ORDERED SEPARATELY FROM DETECTOR ASSEMBLY TO ALLOW FILTER AND INPUT OPTIC SWAPPING.

 

For the latest meter specifications, data sheet, installation guide, tech notes and software, visit the ILT1700 page here.

To view additional Photometry light measurement systems, visit our application page here.